Limit of Iron
Step-by-Step Instructions
-
Add citric acid and thioglycolic acid to the sample solution.
-
Make the solution alkaline with stronger ammonia water.
-
Wait 5 minutes and compare the pink color to a standard iron solution.
Acceptance Criteria
-
Any pink color in the test solution is not more intense than the standard (NMT 2 ppm).